Seamless Brilliance
The TECNIS Symfony® IOL mitigates the effects of presbyopia so you can empower your patients to see clearly across the full range of vision.
See the Passion in Each Patient.
For Each Passion, There’s a TECNIS® IOL. Leading Innovation. High-Quality Vision. Exceptional Satisfaction.
Delivering excellent uncorrected visual acuity (VA) at all distances,1 the TECNIS Symfony® IOL allows your patients to see clearly across a full range of vision.
Defocus Curve1
6-Month Data, Bilaterally Implanted Subjects
Rather than splitting light into distinct focal points, the TECNIS Symfony® IOL elongates focus, resulting in an increased depth of field. This empowers you to deliver a full range of continuous vision while maintaining high image contrast.
Monocular Contrast Sensitivity at 6 Months1
Mesopic without Glare
Enhanced Functionality
Deliver the visual quality that allows your patients with and without astigmatism to see clearly in any lighting condition.
The TECNIS Symfony® IOL delivers 20/20 or better vision through 1.5 D of astigmatism.1
85% of patients wear glasses none or a little bit of the time.1
By combining a unique diffractive pattern with achromatic technology, the TECNIS Symfony® IOL delivers outstanding visual acuity over a continuous range of distances while maintaining high contrast performance.
Degree of Difficulty with Visual Symptoms1
Percentage of Eyes, Non-Directed Reports, 6 Months Postoperative
Low incidence of visual symptoms, including halos, glare and starbursts1
Long-Term Sustainability
Unlike another leading IOL, TECNIS® IOL material is not associated with glistenings,3-7 which can inhibit your patients’ vision. Glistenings cause light scatter, which can result in a reduction in image contrast.4-8
Specifications
TECNIS Symfony® IOL (ZXR00) |
||
OPTIC CHARACTERISTICS |
||
Powers: |
+5.0 D to 34.0 D in 0.5 diopter increments |
|
Diameter: |
6.0 mm |
|
Center Thickness: |
0.7 mm (20.0 D) |
|
Shape: |
Biconvex, wavefront-designed anterior aspheric surface, posterior achromatic diffractive surface designed to reduce chromatic aberration for enhanced image contrast and echelette feature to extend the range of vision. |
|
Material: |
UV-blocking hydrophobic acrylic |
|
Refractive Index: |
1.47 at 35° C |
|
Edge Design: |
ProTEC frosted, continuous 360° posterior square edge |
|
BIOMETRY* |
CONTACT ULTRASOUND |
OPTICAL |
A-Constant: |
118.8 |
119.3 |
Theoretical AC Depth: |
5.40 mm |
5.7 mm |
Surgeon Factor:9 |
1.68 mm |
1.96 mm |
HAPTIC CHARACTERISTICS |
||
Overall Length: |
13.0 mm |
|
Thickness: |
0.46 mm |
|
Style: |
C |
|
Material: |
Soft, foldable, UV-blocking hydrophobic acrylic |
|
Design: |
TRI-FIX, haptics offset from optic, one-piece lens |
IOL INSERTION
Recommended insertion instruments: