Full, Brilliant Vision
The TECNIS® CL IOL delivers ultraviolet-absorbing posterior technology and corrects for corneal spherical aberrations — all in a foldable silicone design.
Three-piece silicone TECNIS® IOLs offer:
Best Case Visual Acuity Results2
AGE |
N |
Percent of “Best Case” Patients with Visual Acuity of 20/40 or Better |
|
TECNIS® CL 3-Piece IOL |
FDA Grid |
||
<60 |
11 |
100% |
96.9% |
60–69 |
73 |
100% |
93.8% |
70–79 |
139 |
98% |
94.9% |
> 79 |
52 |
98% |
87.9% |
Total |
275 |
99% |
94.0% |
Specifications
TECNIS® CL Silicone IOL (Z9002)1 |
||
OPTIC CHARACTERISTICS |
||
Powers: |
+6.0 D to 30.0 D in 0.5 diopter increments |
|
Diameter: |
6.0 mm |
|
Shape: |
Biconvex, anterior aspheric surface, OptiEdge Design |
|
Material: |
UV-blocking SLM-2 Silicone |
|
Refractive Index: |
1.46 at 35° C |
|
CONSTANTS |
Nominal (theoretically calculated) |
|
A-Constant: |
118.7 |
|
AC Depth: |
5.40 mm |
|
Surgeon Factor: |
1.62 mm |
|
HAPTIC CHARACTERISTICS |
||
Overall Length: |
13.0 mm |
|
Style: |
Modified C |
|
Material: |
60% Blue-core Polymethylmethacrylate (PMMA) Monofilament |
|
Angle: |
10˚ |
|
Average Weight in Air: |
19.3 mg |
IOL INSERTION
Recommended insertion instruments: